AODA Explosion-proof Electronic Analytical balance
explosion-proof balance is designed in accordance with GB3836.1-2010 Electrical Equipment for Explosive Gas Environments...
- Fast Delievery
- Quality Assurance
- 24/7 Customer Service
AODA Explosion-proof Electronic Analytical balance
E0652 explosion-proof balance is designed in accordance with GB3836.1-2010 "Electrical Equipment for Explosive Gas Environments Part 4: Intrinsically Safe "i"" in combination with actual use.
It is your ideal choice for measurement in high-risk occasions such as flammable and explosive environments.
Explosion-proof grade: Ex ib IIC T4 Gb;
Explosion-proof certificate number: GYB22.2780X
● Range/division value: 100g/0.1mg, 120g/0.1mg, 200g/0.1mg, 220g/0.1mg, 300g/0.1mg,
● The bottom plate of the shell adopts the integral die-casting aluminum mold one-piece molding process, the upper cover adopts thickened 304 stainless
steel plate, laser nitrogen cutting, high-precision bending molding, cold welding machine welding surface wire drawing treatment process (combination of
machinery and manual), anti-interference, anti-static!
● Weighing pan made of stainless steel, fully transparent glass windshield.
● Fully automatic fault detection, overload protection
● Automatic external weight calibration, full range peeling
● Fast weighing response, reliable accuracy, simple operation, easy maintenance
● Unit conversion, counting function (optional)
● High-definition LCD display, clear and intuitive display
● Adopt external explosion-proof power supply AE0631 for 24-hour uninterrupted power supply,
● Ultra-low power consumption, ultra-long standby function
Series | E0652 explosion-proof Electronic analytical balance | ||||
Capacity | 100g | 120g | 200g | 220g | 300g |
Mini reading | 0.0001g | ||||
Repeatability error | ≦±0.003g | ||||
Linearity error | ≦±0.003g | ||||
Eccentric load error | ≦±0.003g | ||||
Discrimination | 0.0014g | ||||
Explosion-proof grade | Ex ib IIC T4 Gb | ||||
Stabilization time | ≦5s | ||||
Ambient temperature | 0℃~45℃ | ||||
Ambient humidity | 20%RH~60%RH | ||||
Power supply | Explosion-proof power supply AE0631 | ||||
Calibration method | External calibration method: manual loading, automatic calibration | ||||
Tare range | Full range | ||||
Display | LCD, white backlight | ||||
Scale pan size | φ90mm | ||||
Dimensions | 31*21*35cm | ||||
RS232 | Optional | ||||
Applicable scope | Applicable to gas environment, not applicable to dust environment |