Mercury-free pressure differential BOD analyzer

Mercury-free pressure differential BOD analyzer

BOD-573 BOD analyzer works with Mercury-free differential pressure sensing method,can measure 6 samples at the same time....

  • Fast Delievery
  • Quality Assurance
  • 24/7 Customer Service

Mercury-free pressure differential BOD analyzer


BOD-573 BOD analyzer utilizes a mercury-free differential pressure sensing method. Microorganisms in a sealed culture flask undergo aerobic respiration, consuming oxygen and producing carbon dioxide. 

Sodium hydroxide absorbs the carbon dioxide, causing a decrease in pressure within the flask. A pressure sensor monitors this pressure change and converts it into BOD (Body Demand) value.


Intelligent Detection

Independent BOD testing terminal; the BOD analyzer base allows for sample addition to the measurement system at any time, and can simultaneously measure up to 6 samples.

Sample incubation time is settable from 1 to 30 days.

Temperature compensation delay; supports automatic operation after the sample temperature reaches the incubation constant temperature.

The system automatically completes the measurement and automatically shuts down after the measurement is finished, requiring no supervision.


Data Management

Automatically saves process data and supports viewing the most recent set of detailed data.

Test data requires no conversion and can be automatically displayed; test results are stored.

Supports viewing and deleting historical data.

Supports storing 10 years of BOD5 test data.


Power Management

BOD Tester Base, power connector for the incubator

BOD Test Terminal, powered by a high-capacity lithium battery

Automatic sleep and wake-up functions

Backlight brightness adjustment supported


Reference Standard 

HJ 505-2009 Water Quality—Determination of Five-Day Biochemical Oxygen Demand (BOD5)—Dilution and Inoculation Method


Specifications.jpg

Model

BOD-573 BOD analyzer

Measuring range

BOD 0~4000mg/L

Measuring error

Meets water quality BOD5 accuracy testing standards

((Glucose-Glutamic Acid Standard Solution BOD5 = 180~230 mg/L))

Test principle

Mercury-free differential pressure method

Measurement quantity

≤6 pcs

Measuring day

1~30 days

Result Recording Frequency

2 hours/time

Data storage

BOD 5 results for 10years

Culture bottle volume

580mL

Stirring

Magnetic stirrer

Constant temperature

20±1℃

Power supply

DC15V/2A ;15W

Dimensions / Weight

292×200×83mm2.5kg

292×200×345mmincluding bottles